When Does Medical Device Usability Testing Need IRB Review?
When planning a human factors (HF) usability study, medical device manufacturers often ask: Does this study need Institutional Review Board (IRB) review?
The answer isn’t always straightforward. The U.S. Food and Drug Administration’s (FDA’s) requirements for IRB oversight depend on whether a study qualifies as a regulated clinical investigation, the activities involved, and the applicable regulatory requirements.
For HF validation studies intended to support an FDA submission, manufacturers should consider whether IRB review is required or request a formal determination early in the planning process. When the requirements are unclear, a formal IRB determination can help clarify the appropriate level of oversight.
But determining whether IRB review is necessary is not the only consideration. Even when formal review is not required, manufacturers should still examine other factors that protect study participants.
Does FDA Require IRB Review for Usability Testing?
FDA regulations, including 21 CFR Part 50 – Protection of Human Subjects and 21 CFR Part 56 – Institutional Review Boards [2, 3], establish requirements for informed consent and Institutional Review Board oversight in certain clinical investigations involving medical devices.
Not every usability study falls under the same requirements. A simulated-use study, where participants interact with a device in a controlled setting rather than during actual patient care, raises different regulatory considerations than a clinical investigation.
FDA’s Applying Human Factors and Usability Engineering to Medical Devices guidance [1] discusses IRB consultation in the context of actual-use testing. It does not establish a blanket IRB requirement for all simulated-use evaluations. When the answer is unclear, obtaining a formal determination resolves the question before testing begins.
Participant Protection Starts Before the Study
When preparing for a usability study, much of the attention naturally goes toward test scenarios, critical tasks, and the study protocol. Yet some of the earliest participant-related decisions happen during recruitment.
Consider the recruitment process as a whole. How is the study described to potential participants? Is compensation presented appropriately? What personal or medical information is collected during screening?
Recruitment Practices Worth a Closer Look
Recruitment Materials
Recruitment materials, including flyers, emails, and social media posts, may be developed by clients, recruitment agencies, or the human factors team. These materials should accurately describe the study without making unsupported claims about the device.
For example, describing an investigational device as “new and improved” or “safer” may suggest benefits that have not been established. When IRB review of recruitment materials is required, those materials should be included in the review process. Even when formal review is not required, manufacturers should review the materials to ensure that the information communicated to potential participants is accurate and appropriate.
Compensation
Payment is often necessary to recruit qualified participants, especially healthcare professionals. However, compensation should not be presented in a way that creates inappropriate pressure to participate. Payment arrangements should also be clear, including what happens if someone withdraws before completing the study.
Sensitive Screening Information
Screening questionnaires may collect medical histories, information about allergies, or previous device experience. How that information is collected, stored, and retained matters, including for individuals who are screened but never participate.
A Practical Example
Imagine a manufacturer recruiting participants for a simulated-use study involving a wearable medical device. The advertisement calls the product a “new and improved solution” and prominently features the compensation amount. Screening asks about medical history and previous skin reactions.
The study appears straightforward, yet each element raises one of the questions above: an implied unproven benefit, compensation that could overshadow the decision to participate, screening data that must be handled carefully, and skin sensitivities that must be planned for. None of these automatically triggers IRB review, but together they show why participant protection involves more than reviewing the study protocol.
When Should Formative Studies Be Considered for IRB Review?
Not all formative usability studies involve the same level of risk. Interviewing healthcare professionals about an early device concept is different from asking participants to handle syringes, apply adhesive products, or interact with devices containing active medication.
Both may be formative evaluations, but the activities involved introduce different considerations. Rather than relying solely on whether a study is labeled formative or validation, manufacturers should examine what participants will actually experience.
The following factors warrant a closer look when deciding whether IRB review is needed:
| Study Consideration | Examples | Potential Concerns |
| Participant-specific considerations | Minors, individuals with impaired decision-making capacity, or participants susceptible to coercion or undue influence | Additional consent considerations and participant protections |
| Sharp objects and needle-containing devices | Syringes, needles, lancets, autoinjectors, and other devices with exposed or concealed needles | Accidental needlestick injuries, puncture wounds, or other injuries from sharp components |
| Skin contact and material sensitivities | Adhesives, electrodes, wearable devices, latex-containing materials | Skin irritation or allergic reactions |
| Active drug exposure | Studies involving administration or potential exposure to active medication | Additional safety and regulatory assessment; potentially a different review pathway |
These considerations are not an FDA checklist for determining whether IRB review is required, but they help identify situations where additional review or participant protections are appropriate.
Safety Considerations in Simulated-Use Studies
A controlled testing environment can reduce certain risks, but it does not eliminate them. An autoinjector without active medication, for example, may still present a needlestick risk. Similarly, applying an adhesive patch or wearable device may cause skin irritation or an allergic reaction.
Depending on the device and study activities, precautions include screening for known allergies, selecting latex-free materials, limiting exposure to needles or other sharp components, and having procedures in place if a participant experiences discomfort or injury.
What About Low-Risk Formative Studies?
Some formative studies involve little more than interviewing users or gathering feedback on early design concepts. Although these activities may present minimal risk, that alone does not determine whether IRB review is required.
When the regulatory status is unclear, seeking a formal determination and documenting the rationale may be appropriate.
Human factors engineering focuses on understanding the people who use medical devices. That same attention should extend to the individuals participating in usability studies.
Not Sure if Your Study Needs IRB Review?
Core IRB, the Institutional Review Board of Core Human Factors, A Rimkus Company, is registered with the U.S. Department of Health and Human Services and specializes in reviewing behavioral (non-drug) research studies.
Core IRB works with manufacturers and human factors teams to determine whether a study requires IRB review and, when needed, conducts that review with a focus on participant safety and research ethics.
Have questions about IRB oversight for your upcoming usability study?
Visit our IRB Services page to learn more or submit an inquiry to our team.
About the Author
Veronica Mendez, MA, PMP, is a Principal Human Factors Consultant at Core Human Factors, A Rimkus Company, and a member of Core IRB, which provides IRB review services. She brings expertise in human factors engineering, medical device usability testing, and participant safety considerations.